You are here

CMC database

Field of measurement: photometry and radiometry
Quantity: Luminous intensity
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
11.14 (1 - 2.0E+4) cd (0.8 - 0.8) % 818-MP-C811
Quantity: Illuminance responsivity, tungsten source
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(- - -) A/lx (0.9 - 0.8) %
  • Illuminance 0.1lx to 10 lx
  • Correlated colour temperature 2856 K
818-MP-C811
(- - -) A/lx (0.8 - 0.8) %
  • Illuminance 10 lx to 1000 lx
  • Correlated colour temperature 2856 K
818-MP-C811
(- - -) A/lx (0,8 - 1,2) %
  • Illuminance 1000 lx to 30000 lx
  • Correlated colour temperature 2856 K
818-MP-C811
Quantity: Luminous flux
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(10 - 3.0E+4) lm (0,9 - 1,5) %
  • Correlated colour temperature 2000 K to 3200 K
818-MP-C807
11.15 (10 - 2.0E+4) lm (1 - 1) % 818-MP-C807
Quantity: Illuminance
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0.1 - 10) lx (0,9 - 0,8) %
  • Correlated colour temperature 2000 K to 3200 K
818-MP-C811
(10 - 1000) lx (0,8 - 0,8) %
  • Correlated colour temperature 2000 K to 3200 K
(1000 - 3.0E+4) lx (0,8 - 1,2) %
  • Correlated colour temperature 2000 K to 3200 K
818-MP-C811
11.14 (1 - 5.0E+4) lx (0.8 - 0.8) % 818-MP-C811
Quantity: Transmittance, regular, spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0.65 - 1) (0,0008 - 0,0008)
  • Wavelength 380 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0041 - 0,0041)
  • Wavelength 380 nm to 400 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0013 - 0,0013)
  • Wavelength 400 nm to 700 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0017 - 0,0017)
  • Wavelength 700 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,0015 - 0,0015)
  • Wavelength 380 nm to 400 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,00022 - 0,00022)
  • Wavelength 400 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Absorbance, regular, spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0 - 0.19) (0.0003 - 5.0E-5)
  • Wavelength 380 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0019 - 0.0052)
  • Wavelength 380 nm to 400 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0009 - 0.0019)
  • Wavelength 400 nm to 700 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0011 - 0.0025)
  • Wavelength 700 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0022 - 0.0339)
  • Wavelength 380 nm to 400 nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0003 - 0.0048)
  • Wavelength 400 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.03 - 0.0969)
  • Wavelength 380 nm to 400 nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.0017 - 0.0362)
  • Wavelength 400 nm to 1000 nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Wavelength
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(400 - 1650) nm (2.0E-6 - 2.0E-6)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
11.11 (200 - 1000) nm (0.2 - 0.2) nm 818-MP-C810
11.13 (800 - 1650) nm (4.0E-9 - 4.0E-9) 818-MP-C813
Quantity: Spectral reflectance factor R
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
11.1 ((0 až 100) % pro geometrii: 8º / t a 8º / d - (0 až 100) % pro geometrii: 8º / t a 8º / d) ((0,1 + 0,010R) - (0,1 + 0,010R)) %
  • Wavelength (380 up to 460) nm
818-MP-C802
11.1 ((0 až 100) % pro geometrii: 8º / t a 8º / d - (0 až 100) % pro geometrii: 8º / t a 8º / d) ((0,12 + 0,008R) - (0,12 + 0,008R)) %
  • Wavelength 465 up to 780 nm
818-MP-C802
11.1 (Pro geometrii 0º / 45º - Pro geometrii 0º / 45º) ((0,17 + 0,014R) - (0,17 + 0,014R)) %
  • Wavelength (380 up to 780) nm
818-MP-C802
Quantity: Kolorimetrická souřadnice L*
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
11.2 (2 - 99) (0.47 - 0.47) 818-MP-C802
11.2 (-40 - 65) (0.25 - 0.25) 818-MP-C802
11.2 (-45 - 95) (0.25 - 0.25) 818-MP-C802
11.2 (2 - 99) (0.45 - 0.45) 818-MP-C802
11.2 (-40 - 65) (0.35 - 0.35) 818-MP-C802
11.2 (-45 - 95) (0.94 - 0.94) 818-MP-C802
11.2 (2 - 99) (0.35 - 0.35) 818-MP-C802
11.2 (-40 - 65) (0.3 - 0.3) 818-MP-C802
11.2 (-45 - 95) (0.65 - 0.65) 818-MP-C802
11.2 (-45 - 95) (1.45 - 1.45) 818-MP-C802
11.2 (2 - 99) (1.3 - 1.3) 818-MP-C802
11.2 (-40 - 65) (1.55 - 1.55) 818-MP-C802
Quantity: Colorimetric coordinate Y
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
11.3 (0.3 - 90) % (0.55 - 0.55) % 818-MP-C802
11.3 (0.18 - 0.7) (0.0005 - 0.0005) 818-MP-C802
11.3 (0.1 - 0.5) (0.0005 - 0.0005) 818-MP-C802
11.3 (0.3 - 90) % (0.3 - 0.3) % 818-MP-C802
11.3 (0.18 - 0.7) (0.001 - 0.001) 818-MP-C802
11.3 (0.1 - 0.5) (0.0005 - 0.0005) 818-MP-C802