You are here

CMC database

Field of measurement: photometry and radiometry
Quantity: Transmittance, regular spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0.65 - 1) (0,0008 - 0,0008)
  • Wavelength (380 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0041 - 0,0041)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0013 - 0,0013)
  • Wavelength (400 up to 700) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.3 - 0.65) (0,0017 - 0,0017)
  • Wavelength (700 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,0015 - 0,0015)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.02 - 0.3) (0,00022 - 0,00022)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Absorbance, regular, spectral
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(0 - 0.19) (0.0003 - 5.0E-5)
  • Wavelength (380 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0019 - 0.0052)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0009 - 0.0019)
  • Wavelength (400 up to 700) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.19 - 0.52) (0.0011 - 0.0025)
  • Wavelength (700 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0022 - 0.0339)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(0.52 - 1.7) (0.0003 - 0.0048)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.03 - 0.0969)
  • Wavelength (380 up to 400) nm
  • Bandwidth < 3 nm
818-MP-C810
(1.7 - 3) (0.0017 - 0.0362)
  • Wavelength (400 up to 1000) nm
  • Bandwidth < 3 nm
818-MP-C810
Quantity: Wavelength
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
(400 - 1650) nm (2.0E-6 - 2.0E-6) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
(400 - 1650) nm (1.0E-7 - 1.0E-7) (bezrozměrný)
  • CW Single frequency, Power level ˃ 0,05 mW
818-MP-C813
11 (200 - 1700) nm (0.2 - 0.2) nm 818-MP-C810
11 (200 - 1700) nm (0.2 - 0.2) nm 818-MP-C810
16 (800 - 1650) nm (4.0E-9 - 4.0E-9) nm 818-MP-C813
22 (365 - 923) nm (0.4 - 0.4) nm 818-MP-C806
Quantity: Spectral reflectance factor R
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
1 (0 - 100) % ((0,1 + 0,010R) - (0,1 + 0,010R)) % (abs.)
  • Wavelength (380 up to 460) nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,12 + 0,008R) - (0,12 + 0,008R)) % (abs.)
  • Wavelength 465 up to 780 nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,17 + 0,014R) - (0,17 + 0,014R)) % (abs.)
  • Wavelength (380 up to 780) nm
  • geometry 0º / 45º
818-MP-C802
1 (0 - 100) % ((0,1 + 0,010R) - (0,1 + 0,010R)) % (abs.)
  • Wavelength (380 up to 460) nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,12 + 0,008R) - (0,12 + 0,008R)) % (abs.)
  • Wavelength 465 up to 780 nm
  • geometry 8º / t, 8º / d
818-MP-C802
1 (0 - 100) % ((0,17 + 0,014R) - (0,17 + 0,014R)) % (abs.)
  • Wavelength (380 up to 780) nm
  • geometry 0º / 45º
818-MP-C802
Quantity: Surface colour, colorimetric coordinate L*
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
2 (2 - 99) (0.35 - 0.35) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
2 (-110 - 110) (0.25 - 0.25) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
2 (-110 - 110) (0.3 - 0.3) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
2 (2 - 99) (0.4 - 0.4) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
Quantity: Surface colour, colorimetric coordinate Y
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
3 (0.3 - 100) (0.25 - 0.25) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
3 (0.002 - 0.7) (0.0005 - 0.0005) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
3 (0.002 - 0.8) (0.0005 - 0.0005) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
3 (0.3 - 100) (0.35 - 0.35) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
3 (0.002 - 0.7) (0.0007 - 0.0007) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
3 (0.002 - 0.8) (0.0007 - 0.0007) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
Quantity: Surface colour, colorimetric coordinate L
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
4 (0.002 - 0.6) (0.0005 - 0.0005) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
4 (2 - 99) (0.35 - 0.35) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
4 (2 - 99) (0.4 - 0.35) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
4 (0.002 - 0.6) (0.0008 - 0.0008) % (abs.)
  • geometry 8º / t, 8º / d, 0º / 45º
818-MP-C802
Quantity: Gloss
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
5 (0.1 - 150) GU (1.8 - 1.8) GU
  • geometry 20º, 60º, 85º
818-MP-C808 ISO 2813
Quantity: Irradiance
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
7 (0.0001 - 1000) mW/cm2 (5 - 5) %
  • Wavelength (230 up to 470) nm
818-MP-C801
Quantity: Correlated colour temperature
Internal identifier Range Calibration and measurement capabilities Procedure Note to procedure
8 (2000 - 1.0E+4) K (10 - 85) K 818-MP-C801
Quantity: Perpendicular spectral transmittance
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
9 (0.1 - 1) (0.0012 - 0.0012)
  • Wavelength (200 up to 380) nm
818-MP-C810
9 (0.001 - 0.1) (0.0001 - 0.0001)
  • Wavelength (200 up to 380) nm
818-MP-C810
9 (0.6 - 1) (0.0008 - 0.0008)
  • Wavelength (380 up to 1000) nm
818-MP-C810
9 (0.3 - 0.6) (0.0041 - 0.0041)
  • Wavelength (380 up to 400) nm
818-MP-C810
9 (0.3 - 0.6) (0.00056 - 0.00056)
  • Wavelength (400 up to 700) nm
818-MP-C810
9 (0.3 - 0.6) (0.0017 - 0.0017)
  • Wavelength (700 up to 1000) nm
818-MP-C810

Pages