You are here

CMC database

Field of measurement: thermometry and humidity
Quantity: teplota
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
1012-74 (0.01 - 961.8) °C 20 mK
  • temperature controlled 3-zone furnace, furnace, heat pipe and thermostat 
112-MP-C001
(-196 - -196) °C 35 mK
  • boiling point of liquid nitrogen Bath at atmospheric pressure 
133-MP-C002
1012-240 (0.01 - 0.01) °C 4 mK
  • temperature controlled cryostat 
133-MP-C002
1012-250 (0 - 156) °C 14 mK
  • temperature controlled cryostat and water bath  
133-MP-C002
1012-260 (156 - 300) °C 22 mK
  • temperature controlled water bath and oil bath 
133-MP-C002
1012-270 (300 - 420) °C 30 mK
  • temperature controlled oil bath and salt bath 
133-MP-C002
1012-280 (420 - 550) °C 37 mK
  • temperature controlled salt bath 
133-MP-C002
1012-291 (663 - 800) °C 60 mK
  • temperature controlled heat pipe 
133-MP-C002
1012-251 (-30 - 90) °C 10 mK
  • temperature controlled cryostat and water bath  
133-MP-C002
1012-231 (-80 - 0) °C 10 mK
  • temperature controlled cryostat 
133-MP-C002
1012-251 (-80 - 156) °C 14 mK
  • temperature controlled cryostat, water bath and oil bath 
133-MP-C002
1012-261 (156 - 300) °C 22 mK
  • temperature controlled water bath and oil bath 
133-MP-C002
1012-271 (300 - 420) °C 30 mK
  • temperature controlled oil bath and salt bath 
133-MP-C002
1012-281 (420 - 550) °C 37 mK
  • temperature controlled salt bath 
133-MP-C002
1012-291 (550 - 663) °C 44 mK
  • temperature controlled heat pipe 
133-MP-C002
1012-300 (0 - 220) °C 0.4 K
  • temperature controlled water bath and oil bath 
133-MP-C003
1012-310 (220 - 550) °C 0.5 K
  • temperature controlled oil bath and salt bath 
133-MP-C003
1012-320 (550 - 1100) °C 0.8 K
  • temperature controlled salt bath and high temperature furnace 
133-MP-C003
1012-330 (1100 - 1200) °C 1.1 K 133-MP-C003
1012-335 (-196 - -196) °C 0.3 K
  • liquid nitrogen Bath at atmospheric pressure 
133-MP-C003
1012-335 (1100 - 1554) °C 2.2 K
  • high temperature controled furnace 
133-MP-C003
1012-360 (193.2 - 493.1) K 0.4 K
  • temperature controlled cryostat, water bath and oil bath 
133-MP-C003
1012-370 (220 - 550) °C 0.65 K
  • temperature controlled oil bath and salt bath 
133-MP-C003
1012-380 (550 - 1100) °C 1.2 K
  • temperature controlled salt bath and high temperature furnace 
133-MP-C003
1012-385 (1100 - 1400) °C 1.5 K
  • three zone horizontal furnace 
133-MP-C003
1012-301 (419.5 - 419.5) °C 0.12 K
  • three zone horizontal furnace 
112-MP-C002
1012-230 (-80 - 0) °C 10 mK
  • temperature controlled cryostat 
133-MP-C002
(0 - 0) °C 0.01 °C
  • temperature controlled heat pipe 
133-MP-C002
1012-290 (550 - 663) °C 44 mK
  • temperature controlled heat pipe 
133-MP-C002
(-80 - 220) °C 0.4 °C
  • temperatue controlled cryostat, water and oil bath 
133-MP-C003
(0 - 220) °C 0.4 K
  • temperature controlled water bath and oil bath 
133-MP-C003
(220 - 550) °C 0.5 K
  • temperature controlled oil bath and salt bath 
133-MP-C003
(550 - 1100) °C 0.8 K
  • temperature controlled salt bath and high temperature furnace 
133-MP-C003
7 (-189.3 - -189.3) °C 0.9 mK
  • the fixed points of ITS-90 - triple point Ar
112-MP-C001
Field of measurement: electricity and magnetism
Quantity: electromagnetic fields above 50 kHz: electric field strength
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EMF11 (0.01 - 200) V/m (0,4 - 1) dB
  • frequency 50 kHz to 18 GHz
113-MP-C006
Quantity: electromagnetic fields above 50 kHz: power flux density
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
EMF13 (0.1 - 10) mW/cm2 (1 - 1) dB
  • frequency 2450 MHz
113-MP-C006
Quantity: RF power: absolute power in coaxial
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF01 (5.0E-7 - 20) W (4 - 65) mW/W
  • frequency 0 GHz to 18 GHz
  • connector 50 Ω
113-MP-C010
Quantity: RF power: absolute power in waveguide
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF02 (5.0E-7 - 1) W (15 - 50) mW/W
  • frequency 8,2 GHz to 18 GHz
  • connector R100, R140
113-MP-C010
Quantity: RF power: calibration factor in coaxial line
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF03 (0.7 - 1,1) (0,004 - 0,015)
  • frequency 0 GHz to 18 GHz
113-MP-C010
Quantity: Scalar RF reflection coefficient: in waveguide
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF08 (0 - 1) (0.02 - 0,02)
  • frequency 26,5 GHz to 40 GHz
  • connector R320
113-MP-C010
Quantity: Scalar RF attenuation: in coaxial line
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF09 (0 - 110) dB ( 0,02 - 0,57) dB
  • frequency 0,001 GHz to 18 GHz
  • connector 50 Ω
113-MP-C010
Quantity: Scalar RF attenuation: in waveguide
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF10 (0 - 110) dB (0,07 - 0,57) dB
  • frequency 8 GHz to 40 GHz
  • connector R100, R140, R220, R320
113-MP-C010
Quantity: Scattering parameters: reflection coefficient (Sii) in coaxial line, magnitude
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF13a (0 - 1) (0,002 - 0,038)
  • frequency 0.1 MHz to 18 GHz
  • connector N-type, PC-7, PC-3.5
113-MP-C009
Quantity: Scattering parameters: reflection coefficient (Sii) in coaxial line, phase
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF13b (-180 - 180) ° (0,11 - 180) °
  • frequency 0.1 MHz to 18 GHz
  • connector N-type, PC-7, PC-3.5
113-MP-C009
Quantity: Scattering parameters: transmission coefficient (Sij) in coaxial line, magnitude
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF15a (0.0003 - 1) (0,00003 - 0,003)
  • frequency 0.010 GHz to 18 GHz
  • connector N-type, PC-7, PC-3.5
113-MP-C009
Quantity: Scattering parameters: transmission coefficient (Sij) in coaxial line, phase
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF15b (-180 - 180) ° (0,59 - 6) °
  • frequency 0.010 GHz to 18 GHz
  • connector N-type, PC-7, PC-3.5
113-MP-C009
Quantity: Scattering parameters: effective source match
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF17 (0 - 1) (0,008 - 0,03)
  • frequency 0.1 MHz to 18 GHz
  • connector 50 W, 75 W
113-MP-C009
Quantity: antenna properties: antenna gain
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF25 (0 - 28) dB (0.25 - 0,25) dB
  • frequency 1 GHz to 18 GHz
113-MP-C010
Quantity: RF voltage and current: RF voltage sources
Internal identifier Range Calibration and measurement capabilities Conditions of measurement Procedure Note to procedure
RF33 (0.1 - 7) V (4 - 11) mV/V
  • frequency 0 GHz to 2 GHz
  • connector 50 Ω
113-MP-C010, 113-MP-C011
RF33 (0.1 - 7) V (4 - 11) mV/V
  • frequency 0 GHz to 2 GHz
  • connector 50 Ω
113-MP-C010, 113-MP-C011

Pages