You are here

View CMC

Field of measurement

Interní identifikátor

DCV01

Quantity

Instrument or artifact

Calibration and measurement capabilities

Range Calibration and measurement capabilities
1 - 10 V 0,06 - 0,1 µV/V

Technical competence

Guarantee of service

Measurement conditions

fixed voltage 1 V, 1,018 V, 10 V

Method

Number of procedure Title of procedure
Calibration of DC Voltage Zener Reference Standards