Czech metrology institute |
![]() | |
| Official notice board Contacts Problem solutions Public procurement FAQ Events |
EG REFERENCE SOURCES
DESCRIPTION
A weighed amount of the standard solution is dropped on the disc of filter
paper in the polystyrene capsule. The capsule is sealed up, when dried. The capsule
and the reflector layer of common NaI(Tl) scintillators provide a sufficient
filtering of b radiation of
relevant radionuclide. For 144 Ce this filtration is not sufficient
and for the types EG 1 and EG 3 layer of minimum 3,2 mm Al between the standard
and the detector is necessary.
APPLICATION
Energy and efficiency calibration of scintillation spectrometers and counters
of g and X radiation.
They can serve as reference sources for relative measurements. Their activity
is chosen so that: standard EG 1 in closed geometry with the NaI(Tl) 38 x 25 mm
; standard EG 2 inserted to the well of NaI(TI) 45 x 50 mm and standard EG 3
located 10 cm from the forehead of NaI(Tl) 38 x 25 mm gives approx. 1700 counts
per second for energy higher than 30 keV.
MEASUREMENT
The activity of individual standards is calculated from the mass of the
standard solution and is checked by relative measurements of g photon flux. The specific activity is
determined by absolute measurement
using 4p b-g , 4p a-g or 4p X- g coincidence method or 4p proportional counter.
|
Nuclide |
Half life days |
Type |
Activity kBq |
Energy of photons g, keV |
Yield of photons, % |
Uncertainty, % |
Code |
|
22 Na |
950 |
EG 1 |
5 |
511,0 1275,55 |
18,66 99,94 |
1,0 |
NAG - 1 |
|
EG 2 |
3 |
NAG - 2 |
|||||
|
EG 3 |
100 |
NAG - 3 |
|||||
|
54 Mn |
312,22 |
EG 1 |
13 |
846,76 |
99,92 |
0,7 |
MNG - 1 |
|
EG 2 |
6 |
MNG - 2 |
|||||
|
EG 3 |
300 |
MNG - 3 |
|||||
|
57 Co |
271,26 |
EG 1 |
6 |
122,06 136,46 |
85,45 10,77 |
1,0 |
CTG - 1 |
|
EG 2 |
2 |
CTG - 2 |
|||||
|
EG 3 |
150 |
CTG - 3 |
|||||
|
60 Co |
1925,4 |
EG 1 |
10 |
1173,21 1332,47 |
99,865 99,981 |
0,7 |
COG - 1 |
|
EG 2 |
4 |
COG - 2 |
|||||
|
EG 3 |
200 |
COG - 3 |
|||||
|
65 Zn |
243,9 |
EG 1 |
40 |
1115,52 |
50,75 |
1,6 |
ZNG - 1 |
|
EG 2 |
18 |
ZNG - 2 |
|||||
|
EG 3 |
800 |
ZNG - 3 |
|||||
|
88 Y |
106,6 |
EG 1 |
8 |
898,021 1836,030 |
93,52 99,36 |
1,2 |
YWG - 1 |
|
EG 2 |
4 |
YWG - 2 |
|||||
|
EG 3 |
200 |
YWG - 3 |
|||||
|
129 I |
5,734.10 9 |
EG 1 |
15 |
39,58 29 - 33 X K |
7,46 > 70 |
0,7 |
IZG - 1 |
|
EG 2 |
5 |
IZG - 2 |
|||||
|
133 Ba |
3897 |
EG 1 |
3 |
80,989 302,795 355,95 |
34,2 18,46 62,22 |
0,8 |
BAG - 1 |
|
EG 2 |
2 |
BAG - 2 |
|||||
|
EG 3 |
80 |
BAG - 3 |
|||||
|
137 Cs |
11019 |
EG 1 |
16 |
661,649 |
85,10 |
0,9 |
CSG - 1 |
|
EG 2 |
7 |
CSG - 2 |
|||||
|
EG 3 |
300 |
CSG - 3 |
|||||
|
141 Ce |
32,50 |
EG 1 |
10 |
145,444 |
48,43 |
0,9 |
CKG - 1 |
|
EG 2 |
3 |
CKG - 2 |
|||||
|
EG 3 |
250 |
CKG - 3 |
|||||
|
144 Ce |
284,4 |
EG 1 |
30 |
133,531 |
11,09 |
1,1 |
CEG - 1 |
|
EG 2 |
8 |
CEG - 2 |
|||||
|
EG 3 |
600 |
CEG - 3 |
|||||
|
152 Eu |
4858 |
EG 1 |
30 |
from 121 to 1538 keV |
depends on energy |
0,8 |
EUG - 1 |
|
EG 2 |
15 |
EUG - 2 |
|||||
|
EG 3 |
450 |
EUG - 3 |
|||||
|
203 Hg |
46,72 |
EG 1 |
8 |
279,19 |
81,49 |
1,1 |
HGG - 1 |
|
EG 2 |
3 |
HGG - 2 |
|||||
|
EG 3 |
200 |
HGG - 3 |
|||||
|
241 Am |
157800 |
EG 1 |
15 |
59,5364 |
35,67 |
0,6 |
AMG - 1 |
|
EG 2 |
5 |
AMG - 2 |
|||||
|
EG 3 |
450 |
AMG - 3 |


Type EG 1 and EG 3


Type EG 2
Annotation:
Uncertainty is an abbreviation for combined standard uncertainty ( P = 68,3 %).
MBSS STANDARDS IN MARINELLI BEAKERS
DESCRIPTION
Marinelli beakers are filled with silicone rubber containing uniformly
distributed radionuclide or mixture of radionuclides. Default density of the
active volume is 0,98 g.cm -3 and mean atomic number approaches water.
The standards are available in 3 types of polypropylene beakers with default
volumes 450, 500 and 1000 ml. Other volumes, nuclides, activities or beakers
are available on request.
APPLICATION
Energy and efficiency calibration of gamma spectrometers.
MEASUREMENTS
The standards are prepared from the standard solutions ER(EB) whose activity is
determined by absolute method. The final sources are checked by measurement on
gamma spectrometer with HPGe detector. Combined standard uncertainty of activity
is approx. 2 %.
|
Type |
Nuclide |
Half life, days |
Activity, kBq |
|
MBSS 1 |
152 Eu |
4858 |
3 |
|
MBSS 2 |
mixture according to IEC 697/81 |
- |
40 |
|
MBSS 3 |
134 Cs |
753 |
* |
|
MBSS 4 |
137 Cs |
11019 |
3 |
|
MBSS 5 |
226 Ra |
584300 |
3 |
|
MBSS 6 |
57 Co |
271,26 |
* |
|
MBSS 7 |
60 Co |
1925,4 |
3 |
|
MBSS 8 |
241 Am |
157800 |
10 |
|
MBSS 9 |
232 Th |
5,15 . 10 12 |
1,5 |
|
MBSS 10 |
153 Gd |
241,6 |
* |
|
MBSS 12 |
133 Ba |
3897 |
* |
|
MBSS 13 |
109 Cd |
462,6 |
20 |
|
MBSS 14 |
210 Pb |
8108 |
* |
|
MBSS 15 |
192 Ir |
74,12 |
* |
|
MBSS 16 |
85 Sr |
64,78 |
* |
|
MBSS 17 |
54 Mn |
312,22 |
* |
|
MBSS 18 |
88 Y |
106,60 |
* |
|
MBSS 19 |
139 Ce |
137,50 |
* |
|
MBSS 20 |
40 K |
4,602 . 10 11 |
1,5 |
* - on request




Type 0530G Type 0540G Type 1040G
Annotation:
Uncertainty is an abbreviation for combined standard uncertainty ( P = 68,3 %).